Master switch on stirrer won't light
1. check power cord and power supply
Item won't work (master switch lighted))
2. Thermal motor circuit breaker triggered by motor overload
Remedy:
Wait about 20 minutes, decrease load applied on shaker plate.
Motor running, output shaft static
3. rubber ring or friction cone are worn out
Remedy:
replace rubber ring or friction cone and adjust press load
RZR 1
Model
Motor power- input/output
Number of gear stages
Speed range
Speed control
Max.torque
Torque at 1000 min
Dimensions of unit (BxHxT)
Dimensions of stay bar (
Weight
Dia. of stirrer shaft (plug in)
Operating temperature range
Max.permissible relative humidity
protection class
Interference suppression standard
Motor protection feature
Permissible operating time period
TROUBLESHOOTING
SPECIFICATIONS
(1/min)
-1
(mm)
xL)
(mm)
(kg)
(mm)
(°C)
DISPOSAL
(W)
35-250 and 280-2200
mechanically stepless, scale with guide
(Ncm)
(Ncm)
71 wide; 250 high; 172 deep
max. 80% (avoid reaching dewpoint)
overheating protection, self-resetting
20
RZR 1
77/18
2
numbers
100
8
13x300
2.5
max.8,0
0-40
IP 20
VDE 0875
100%