Error messages and Trouble shooting
Temperature Error
In order to get accurate test result, perform testing between
1. When the ambient temperature is 32~48℉ (0~9℃) or 106~122℉ (41~50℃) the "
will be blinking (1a), you still can do the test but the test result is only for reference, because the test
result under these ranges of temperature might not be correct. In the memory screen this value will be
flagged with "
" symbol (1b). Repeat the test at an area with temperature between operating range
(50~104℉ or 10~40℃).
2. When you move from the area with temperature outside the operating range of test strip to another area
with temperature inside operating range of test strip, please wait for 30 minutes before you do the test.
3. When the temperature is below 32℉ (0℃) or over 122℉ (50℃),
can not do the test and the "
" symbol will blink with thermograph symbol displaying on screen in
this condition. Please move the meter to environment with temperature between 50~104℉ (10~40℃)
and repeat the test after 30 minutes.
46
50~104℉ 10~40℃
(
).
1a
" warning symbol
3
Rightest ®
Blood Glucose Meter GM550
Error messages and Trouble shooting
Sampling Error
Please don't apply the blood drop to the sample entry of the strip before the
Meter GM550 displays "
". If you do so, the meter will display "
(if volume is turned on). Please discard the test strip and repeat the test with a new
Glucose Test Strip GS550.
Meter Malfunction
®
If the
Rightest
Blood Glucose Meter GM550 can not be started, please follow the steps below:
1. Open the battery cover, remove the batteries.
2. Wait for 5 minutes and insert the batteries as described "Meter activation and Battery change"
on page 16~17.
®
The
Rightest
Blood Glucose Meter GM550 should be working normally after finishing above steps. If not, please
contact your local Bionime customer service.
®
Rightest
Blood Glucose
" and "
" accompanied by beeps
®
Rightest
Blood
47