To avoid RCD circuit-breaker tripping, a test must be carried
out between L and N for approx. 5s.
Immediately afterwards, voltage testing between L and PE can
be carried out without the RCD switch tripping.

4.6 Continuity Test / Diode Test

Ensure that the UUT is not live b y disconnecting from
all live circuits.
The test voltage polarity at the test probe is negative (-).
Open Polaris 3 plus Voltage Tester as described on page
26.
Check that UUT is not live, by applying double-pole vol-
tage test.
Connect both test probes to UUT.
An acoustic signal
splayed to indicate continuity.

4.6 Rotary Field Indication

Polaris 3 plus Voltage Testers equirred with a two-pole Rota-
ry Field Indicator.
Operator's safety measures must be met, as described
in paragraph 2.0.
The Rotary Field Indicator is always active. The symbol
R
or
fields can only be determined in three-phase systems. The in-
strument indicates the voltage between the two external con-
ductors.
Open Polaris 3 plus Voltage Tester as described on page
26.
Connect instrument test probe with supposed phase L2
and hand test probe with supposed phase L1. Touch
contact electrode (5).
30
is audible, the continuity symbol is di-
L
is always displayed. The direction of rotary