Product Description; Fields Of Application; Technical Data; Installation - 3M ESPE CapMix Instrucciones De Servicio

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 19
14. Only appropriately labeled original 3M ESPE replacement
parts may be used in the repair and maintenance of this
unit. Any damage resulting from the use of third-party
replacement parts voids the liability of 3M.
15. Do not attempt to operate this unit by bypassing the
protective mechanisms (e.g. by turning off the automatic
shut-off switch on the cover of the unit). Violation of the
safety regulations may cause serious injury and voids
all warranties and guarantees.
16. Do not expose the device to solvents, flammable liquids,
and strong heat sources as these may damage the
plastic housing.
17. Cleaning the unit, ensure that no cleaning agent enters
the device since this may lead to electric short-circuiting
or hazardous malfunctions.
18. Pay attention to capsule leak tightness when mixing.
Capsules that are not leak tight contaminate the device
with mercury. Contaminated devices cannot be repaired
and must be disposed of as contaminated waste
according to local laws.
See the safety data sheet for the amalgam used for
first aid, measures to take in case of unintended release,
handling, storage, exposure limitation, and personal
protective clothing.

2. Product Description

The CapMix unit is a mixing device for dental filling
materials and cements in capsules. Desired mixing times
may be programmed into the mixing programs "P1"-"P4".
Moreover, two pre-set mixing times, "10 s" and "15 s", can
be selected.
Operating Instructions should not be discarded for the
duration of product use.

3. Fields of Application

Dental materials packaged in capsules with a max. capsule
weight of 7 g.
Capsules with a weight of more than 7 g must not be used!

4. Technical Data

Drive unit:
Operating mode:
Controller:
Capsule holder:
2
Ball-bearing motor with integrated
self-resetting overheat cut-off
Short-term operation,
1 min On/7 min Pause
Microprocessor-controlled,
high-precision mixing time control
by quartz timing mechanism
Self-locking
Capsules:
Housing:
Mixing speed:
Operating temperature: 18°C...40°C/64°F...104°F
Relative humidity:
Main voltage:
Power requirements:
Fuses:
Dimensions:
Weight:
Continous sound
pressure level:
Year of manufacture:

5. Installation

Prior to the initial start-up, remove the transport
safety guard.
Connect the CapMix unit to the power supply and switch on
the unit with the power switch on the back of the device.
The display shows the mixing time of program 1.
The following basic settings are preset at the factory:
Program 1 (key "P1"):
Program 2 (key "P2"):
Program 3 (key "P3"):
Program 4 (key "P4"):
Fixed time (key "10 s"): 10 sec mixing
Fixed time (key "15 s"): 15 sec mixing

5.1 Setting the mixing time

Mixing times in the range of 1sec to 39 sec may be selected.
By pressing the keys "+" or "-" the device increases or
decreases, respectively, the mixing time by 1sec. Depressing
and holding the "+" or "-" keys, the device continues to
count up or down, respectively.

5.2 Changing the basic settings

Set the mixing time as described above, then depress one
of the program buttons, "P1"-"P4", until two short signals
are issued. This confirms that the desired settings have
been stored in the selected program.
Diameter max.19 mm,
total length max. 58 mm,
length without tubes max. 38 mm,
weight max. 7 g
Housing and protective cover made
of high impact-resistant plastics
4650 rpm for 7 g-capsules
max. 80% at 31°C/88°F,
max. 50% at 40°C/104°F
100 V, 120 V, 230 V, 50/60 Hz,
for preset voltage refer to the
rating plate
330 VA
230 V: 2 A time-lag fuse
120 V: 3.2 A time-lag fuse
(W x H x D) 195 x 160 x 195 mm
3750 g
>70 dB(A)
see rating plate
5 sec mixing
7 sec mixing
9 sec mixing
12 sec mixing

Publicidad

Tabla de contenido
loading

Tabla de contenido