somnibel
User's Manual
Somnibel
Do not forget to unclip the fastening adhesive from
after each use. This will expand battery life.
Somnibel
If
indicates low battery after clipping the fastening adhesive, connect the device to mains before using
it for assuring a complete therapy during the whole night.

3.1 SOMNIBEL FUNCTIONING

Somnibel
is automatically turned ON when the fastening adhesive is clipped to the device. At the beginning, the
device performs a self-test for checking its correct functioning. The visual indicator on the device will provide
indications of the results of self-test and battery status (see sections VISUAL INDICATIONS AND WARNINGS).
Somnibel
After turning ON,
will wait for 15 minutes before starting the stimulus application so that the patient
can restfully get asleep.
Somnibel
device generates a gentle vibrating stimulus when the supine position is detected for inducing the
patient to change his/her body position. If the patient does not change the position, the device will wait for 30
seconds and then it will increase the stimulus intensity. The intensity will be progressively increased until the
patient changes the position (4 intensity levels).
EN 12
534-800-MUM_Rev.1.06