EMC Technical Information
Table 1
The "CALAJECT™" is intended for use in the electromagnetic environment specified below. The customer or the user
Emissions test
RF emissions
CISPR11
IEC 60601-1-2
EN55011
RF emissions
CISPR11
IEC 60601-1-2
EN55011
Harmonic emissions
IEC 60601-1-2
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 60601-1-2
IEC 61000-3-3
Table 2
The "CALAJECT™" is intended for use in the electromagnetic environment specified below. The customer or the user
Immunity test
Electrostatic discharge
(ESD)
IEC 60601-1-2
IEC61000-4-2
Electrical fast transient/
burst
IEC 60601-1-2
IEC61000-4-4
Surge
IEC 60601-1-2
IEC61000-4-5
Voltage dips, short
interruptions and voltage
variations on power supply
input lines.
IEC 60601-1-2
IEC61000-4-11
Power frequency (50-60Hz)
magnetic field.
IEC 60601-1-2
IEC61000-4-8
10
Electromagnetic emissions
of the "CALAJECT™" should assure that it is used in such an environment.
Compliance
Group 1
Class B
Class A
Complies
Electromagnetic immunity
of the "CALAJECT™" should assure that it is used in such an environment.
IEC 60601-1
Test level
±6KV contact
±8KV air
±15KV air
±2KV for power supply
lines
±1KV for input/output lines
±1KV differential mode
<5% UT for 0.5 cycle
40%UT for 5 cycles
70%UT for 25 cycles
<5%UT for 250 cycles
30A/m
Electromagnetic environment guidance
The device uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
The device is suitable for use in all establishments, including
domestic establishments and those directly connected to the
public low-voltage power supply network that supplies build-
ings used for domestic purposes.
Compliance level
±6KV contact
±8KV air
±15KV air
±2KV for power supply
lines
±1KV for input/output lines
±1KV differential mode
<5% UT for 0.5 cycle
40%UT for 5 cycles
70%UT for 25 cycles
<5%UT for 250 cycles
30A/m
Electromagnetic envi-
ronment guidance
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the rela-
tive humidity should be at
least 40%.
Mains power supply quality
should be that of typical
residential area.
Mains power supply quality
should be that of typical
residential area.
Power frequency magnetic
fields should be that of
typical residential area.