Position the CryoPen/CryoProbe for treatment
The instrument must be held vertically with the tip directed downward.
An angle of less than 45° must be avoided.
Effective application
Before starting treatment take into account age, location,
size and number of lesions, skin type, degree of tanning and season.
OK!
The applicator tip (from where the liquefied N
O flows) should
2
be held 0,5 to 3 mm from the lesion. Further than 5 mm from
Position the CryoPen/
the applicator tip the liquefied N
O will have expanded into gas phase
CryoProbe correctly
2
which is not effective for achieving therapeutic results.
By approaching the lesion with the gas flowing you will observe
the focal point of the jet on the tissue.
The closer you approach the lesion, the more freezing power you apply
to the tissue. It is recommended to keep a greater distance for the
smaller and more delicate procedures.
The amount of liquid N
O that you wish to apply can be controlled
Distance to lesion
2
by moving the applicator quickly towards and away from the tissue.
Circular and crossing movements have the same effect.
15