7. Operation

Activate the main switch (2) on the back
of the device; the functional display LED
(6) lights up in red. The device is now
ready for operation.
Open the polymerisation chamber by
raising the device cover (12) at the han-
dle (11). The opening angle is approx.
55° and is fixed in its final position by
a magnet.
The central point of the surface that is to
be polymerised must always be positioned
on the seating for the model at a distance
of 2-3 cm from the middle LED that has a
black circle round it.
Fixing the model vertically on the pendu-

7.1 Time programmes

The EyeVolution
MAX light-curing
®
device features several pre-programmed
time relays. Use the push-button (4) to
activate time programmes with 1 / 2 / 3 / 5
and 10 minutes.
The following times apply, depending
on the composite material that is being
used and for a maximum layer thickness
of 2 mm:

7.2 Rotary plate

A reflective pendulum plate (7) was
designed for the seating of the model to
ensure optimum polymerisation with as little
shadowing as possible. Starting in a central
position (see point 7.1), the seating of the
model first moves to the right by 90° after
the program is started and then shifts into a
180° pendular motion. The oscillation func-
tion is coupled with the time programmes
lum plate is recommended for exclusively
occlusal modelling (e.g. inlay or onlay).
optimal
start position
Release the cover from its magnetic
anchoring by pulling slightly on the han-
dle and close the polymerisation chamber.
Opaquer: 1:00–2:00 min
Pre-polymerisation: 0:30–1:00 min
Interim polymerisation: 1:00–2:00 min
End polymerisation: 2:00–5:00 min
Start the selected time programme by
pressing the corresponding push-button
(4). The polymerisation process starts
automatically and is indicated as an active
programme by a green LED.
and is automatically activated when the
polymerisation programme starts. The LED
(6) indicates the operating status of the
rotary plate:
Green LED = rotary plate in action
Red LED = rotary function deactivated
Use the push-button (5) to activate or deac-
tivate the rotary plate function at any time,
even during on-going polymerisation pro-
grammes.