Ranges
Input/output ratio
1000 A
4.1/ RÉFERENCE CONDITIONS
- Temperature: 18...28°C
- Humidity rate: 20...75% RH
- Battery voltage: 9 V ± 0.1 V
- Position of conductor: centred on the markings of the clamp
- Magnetic field: earth's DC field
- Absence of external AC magnetic field
- Absence of electric field
- Measurement for a DC current or an AC sinusoidal current £ 65 Hz
- Impedance of the measurement instrument: ³ 1 MW and £ 100 pF
4.2/ OPERATING CONDITIONS
The instrument must be used in the following conditions to satisfy
the safety of the user and the metrological performance:
- Use indoors
- Altitude: £ 2000 m
- Transportation altitude: £ 12 000 m
% RH
9 0
8 0
7 0
6 0
5 0
4 0
3 0
2 0
1 0
0
-50
-40
4/ SPECIFICATIONS
1 mV/A
Storage
S t o c k a g e
-30
-20
-10
0
Measurement extent
A AC rms
0.5...1000
Use
Utilisation
1 0
2 0
3 0
Temperature
12
A peak max
0.5...1400
Reference
Réference
4 0
5 0
6 0
7 0
A DC
0.5...1400
°C
8 0