Proportion of capacity weight for SPAN adjustment
OPERATION
+
+
+
Ensure there is no weight on the pan
Place the calibration weight on the pan (e.g.
3kg)
Calibration completed
Put 6kg weight on the pan
Escape from Calibration mode
OPERATION
+
+
+
+
+
Ensure there is no weight on the pan.
1 - ZERO
2 - NET
DISPLAY 1 2
0.0 0 0
Weighing mode
8 8 8 8 8 8
Keep pressed the
C A L 0 0
- - - - - -
Calibrating zero point.
C A L S P
Display will show "CALSP" and full capacity (eg 6kg) alternately.
6.0 0 0
5.4 0 0
Press
4.8 0 0
Decreased at 10% of capacity rate.
3.0 0 0
Set to 3kg.
- - - - - -
Calibrating Span
3.0 0 0
The balance goes back to weighing mode.
6.0 0 0
1 - ZERO
2 - NET
DISPLAY 1 2
0.0 0 0
Weighing mode
8 8 8 8 8 8
Keep pressed the
C A L 0 0
0.0 0 0
Escape from calibration and go back to weighing mode.
8 8 8 8 8 8
Keep pressed the
C A L 0 0
- - - - -
Calibrating zero point.
C A L S P
0.0 0 0
Escape to weighing mode.
REMARKS
key and press
, , at this moment release the
to set weight used for calibration.
REMARKS
key and press
, at his moment release the
key and press
, at this moment release the
+
+
key
+
+
key.
+
+
key