Reli On PRIME Manual De Instrucciones página 41

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Step 4
Bring meter and test strip to the drop. Test strip will draw up the
solution. The meter will show result in 7 seconds.
NOTE:
Dye added to control solution may stain surfaces or
fabric. Use with care.
Step 5
Compare the result with the range printed on
the test strip bottle you are using. Values may
vary from bottle to bottle. Verify that the result is
within range of the control solution level that you
are using. If the result falls within the range on
the bottle, the meter is working correctly.
Step 6
Remove the test strip. The meter shuts off automatically when the test
strip is removed.
Dispose of the test strip immediately.
06146A
2016-12
Control Solution Range (mg/dL)
(Rango de la solución de control)
80-120
Level 1:
200-300
Level 2:
41

Publicidad

Tabla de contenido
loading

Productos relacionados para Reli On PRIME

Tabla de contenido