Operation
5. Desinfección
6. Enjuague final
7. Inspection
8. Packaging
9. Sterilization
Operating mode
- Sumergir los accesorios en
una solución desinfectante
(bactericida, virucida, fungicida,
tuberculocida y aldehído libre)
según las recomendaciones del
fabricante
- Ver punto # 2
- Después de enjuagar, los
accesorios deben secarse.
- Inspect devices and sort out
those with defects.
- Pack the devices in
"Sterilization pouches".
- Steam sterilization at:
134℃, 2.0bar~2.3bar
(0.20MPa~0.23MPa), 4min.
- Siga las instrucciones y observe las
concentraciones y el tiempo dado por el
fabricante.
- Dirty instruments must be cleaned and
disinfected again.
- Check the validity period of the pouch given
by the manufacturer to determine the shelf life.
- Use packaging which are resistant up to a
temperature of 141°C and in accordance with
EN ISO 11607.
- The accessories (lip hook, file clip and touch
probe) must be sterilized according to the
packaging labeling. - Use fractionated vacuum
or gravity (less preferred) autoclaves (according
to EN 13060, EN 285). - Use validated
sterilization procedure according to ISO 17665-
1 - Respect maintenance procedure of the
autoclave device given by the manufacturer. -
Use only the listed sterilization procedures.
24
Warning