After cleaning/sterilising, the turbine should be rinsed in distilled water and
left to dry.
Do not use objects, such as pens or cotton buds, to clean or dry the
turbine, as this may cause irreparable physical damage.
To replace the turbine hold the unit and the turbine swirl plate facing you,
put the turbine back into the turbine housing then push until the turbine is
fully located, such that the back of the turbine is completely level with the
back of the unit.
The PulmoLife moulding can be wiped clean using a damp cloth or surgical
wipe.
CALIBRATION CHECK
The unit should not require recalibration. However, as a
confirmation of the correct functioning of the unit, the calibration
can be checked using a 3-litre calibration syringe.
To check the calibration hold down the On
and Scrolling Up
keys together, and only release when 'Ch' is displayed.
When the blow symbol
and '0.00' appear, empty the 3-litre syringe
smoothly into the unit.The value will be displayed.
If the unit detects an error the blow symbol and calibration value will flash.
An acceptable calibration check is within +/-3% of the 3-litres entered (so
between 2.91 and 3.09 litres).
PLEASE NOTE: Calibration check failure is unusual but is most often due
to malfunction of the turbine, either because it is physically damaged or
dirty. Clean the turbine (see cleaning) and recheck the calibration. If the
unit still fails calibration check, replace with a new turbine (see accessories
for order number).
SERVICE
This unit requires no routine servicing.