en - Areas of application
32
When all instruments have been arranged, lower the rail to secure
the connectors and click it into place.
The water supply pipe retaining plates are secured in their positions
by the rail.
Water supply pipes which are not being used must be pushed into
the holder behind the water inlet tube so that they are not thrown
about by the force of the water during the cleaning process.
Instruments which cannot be secured by the retaining plates and
rail cannot be reprocessed in this unit.
Routine checks of cleaning performance reliability
Instruments should be dismantled on a sample basis after their last
operative use and machine reprocessing.
To do this cap the Bowden cables on the guide wheels and pull off
and dismantle the distal functional part.
The inner distal cylinder through which the Bowden cables are
passed to the functional end must be checked optically with
magnification and by protein analysis.