Calibration; Charging; Special Functions; Product Information - Quest SD-200 Manual De Usuario

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

12

Calibration

Calibration
Calibrator

Charging

Using the computer's USB port, the SD-200 is charged with a mini-USB cable
plugged into the bottom of the instrument. It will charge in 2.5 to 12 hours when the
battery is fully discharged based on available power.

Special functions

LED's: light-emitting diodes are activated in the LED alert screen. They will flash at the
following levels for SD-200 kit: 70-0715-6546-2. (Note: These are the "U" country code
levels):
Green LED flashes when measurements are between 80 to 85 dB
Yellow LED flashes when measurements are 85 to 105 dB
Red LED flashes when measurements are above 105 dB.
Note: For SD-200-BR kit ; ID: 70-0716-2264-4, with country code "b" the levels include:
Green LED flashes when measurements are between 43 to 80 dB
Yellow LED flashes when measurements are 80 to 85 dB
Red LED flashes when measurements are above 85 dB.

Product information

 Sound Detector Kit, USB Cable, Windscreen (SD-200)

Accessories

 Windscreen (one included ) (TSI ID: 053-686)
 Mini-USB cable (one included) (TSI ID: 053-575)
 056-990 0.5" Calibrator adapter (TSI ID: 056-990)
 Calibrator: AC-300: 114 dB at 1 kHz output (TSI ID: AC-300)
o QC-10 calibrator (obsolete) is also acceptable for calibrating the SD-200 at 114 dB
at 1 kHz output.
 Tripod (TSI ID: 059-045)
Calibrations should be performed at 114 dB at 1kHz.

Publicidad