Safety information for the individual operating phases

2.8.1
Transport
Transport the device only as described in the Operating Instructions (see requirements
regarding transport and setting up in the Operating Instructions, "Document of equal
applicability and required files", page 7).
In order to avoid corrosion on the device and subsequent damage to the device, transport
and store device only in the permissible temperature range and ambient conditions (see
requirements regarding the ambient conditions in the Operating Instructions, "Document
of equal applicability and required files", page 7).
2.8.2

Initial operation

Ensure that this device is only operated by authorized and trained specialists.
If the device has been stored in a cold environment or at high humidity, maintain a drying
or temperature adjustment time of approx. 4 hours (without voltage) at room
temperature before initial operation.
Before connecting the device, check that the supply voltage and frequency are correct at
the point of installation (see electrical data in the Operating Instructions, "Document of
equal applicability and required files", page 7).
The requirements regarding the location site and ambient conditions must be observed
(see requirements regarding the location site and ambient conditions in the Operating
Instructions, "Document of equal applicability and required files", page 7).
2.8.3
Operation
The device may only be operated by authorized and trained technical personnel.
Keep unauthorized persons, such as patients, children and animals, away from the device.
In all cases of doubt regarding the safety of the appliance, switch off the appliance and
take suitable measures to prevent further use.
Prior to connecting power or operation, check the device, the accessories and protective
equipment for any damage.
Do not use damaged, non-functioning equipment or accessories; instead notify your
authorized service partner.
In order to ensure the product safety and warranty services, the device must be exclusively
operated with the original accessories from Ivoclar Vivadent, particularly the original
power cord. The user bears the risk when using non-approved accessories.
To ensure process reliability, Ivoclar Vivadent recommends using only approved materials.
Never bypass the safety equipment of the device or set it out of operation ("Working
areas, potential hazards and safety measures on the device", page 12).
Operate the device only with the build chamber closed.