wtw Turb 430IR Instrucciones Breves De Operación página 9

Turbimetro
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Turb 430 IR
Inserting a 28 mm cell
Aligning the cell
ba75506e01
04/2005
1
2
Insert the cell so that it is positioned on the bottom of the cell
shaft.
The cell is ready to be measured.
3
Align the cell (see below).
1
Clean the cell.
2
Insert the cell.
3
Align the cell:
! Press and hold the <START/ENTER> key.
! Slowly and in small steps rotate the cell by one complete
rotation (by 360 °).
After each step wait for a short time until the displayed
measured value is stable.
! Turn the cell back to the position with the lowest measured
value.
4
Release the <START/ENTER> key.
Measurement starts. The measured value is displayed.