L O w A i r L O S S M At t r e S S r e p L A C e M e n t S y S t e M

6.0 trOuBLeShOOtinG

problem
1.) Power
control unit
is working,
but mattress
replacement
is not inflating.
2.) Patient is
"bottoming
out".
3.) Power
unit does not
operate.
inspection procedure
1.1) Verify air is flowing smoothly through
hoses and mattress manifolds. Inspect
for cuts or cracks.
1.2) Verify hoses are properly connected
to power unit.
1.3) Verify control panel Power is ON.
2.1) Look at Patient Weight setting on
power unit.
2.2) Check for mattress leaks.
2.3) Check Air Filter for dirt/lint.
3.1) Verify power cord is securely
plugged into wall outlet.
3.2) Verify power cord is securely
plugged into power unit.
3.3) Verify power switch on left side
panel in ON position
3.4) Power surge shuts down pump.
3.5) Power unit does not respond to
possible solutions listed in 3.1 thru 3.4.
possible Solutions
1.1) Hose(s) or manifolds may need to
be moved to avoid kinking or obstruc-
tion. If cuts or tears are observed,
hose(s) or manifold may need to be
replaced.
1.2) Attach CPC connectors securely
into place.
1.3) Press Power button on and LED
will illuminate.
2.1) Increase or decrease weight set-
ting using Patient Weight buttons until
adequate pressure setting is achieved.
2.2) Replace with appropriate spare
parts.
2.3) Clean or replace Air Filter.
3.1) Secure plug into wall outlet.
3.2) Secure plug into power unit.
3.3) Turn power switch to ON position.
3.4) A power surge may temporarily
overload the circuitry. Turn unit off for
several seconds. Check the fuse for
damage. Re-try turning unit on with
normal operating procedures.
3.5) Call customer service at 1-800-333-
4000 or 1-800-263-3390 in Canada.