Message « remove mouthpiece » appears several
times in a row.
If the mouthpiece has been inserted while the cleaning
phase is in progress, then the OCIGO will ask you to
remove the mouthpiece and resume its cleaning cycle
from the beginning. Please wait until the « cleaning »
message disappears and the « insert » « mouthpiece »
message is displayed.
Messages « T < 5°C » or « T > 45°C » followed by OCIGO
turning off.
OCIGO detects a temperature range below 5°C or above
45°C. Therefore it will not work. Please wait until the
temperature is above 5°C or below 45°C before turning
it back on.
Message « T 10°C - 40°C » when starting up the
OCIGO.
The device detects a temperature range of between 5°C
and 10°C or between 40°C and 45°C. You can perform
a measurement, but it will not be reliable because it is
outside the operating range. OCIGO can only operate
correctly when the ambient temperature is between 10°C
and 40°C.
The levels displayed by the OCIGO seem inconsistent
To take a correct measurement, it is essential to comply
with the following rules:
- Use a new mouthpiece for each test;
- Wait 20 minutes after your last glass of alcohol or 5
minutes after smoking;