TSI Alnor 9880 Manual Del Usuario página 21

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Example No. 1
You want to measure the actual velocity in a duct.
The air temperature in the duct is 55°F and the
pressure is 14.24 psia. You take a measurement and
the display reads 1200 fpm.
=
Act Vel
.
.
Example No. 2
You need to measure the actual velocity in a
plenum. The air pressure is 99.4 kPa and the
temperature is 27°C. The display reading on the
Model 9880 is 2.30 m/s.
=
Act Vel
.
.
460 + 55
1200
 
460 70
273 + 27
2 30
.
 
+
273 211
14 7
.
 
+
14.24
1014
.
 
.
99.4
=
1203 7
.
ft
=
2 39
.
m s
/
/ min
19

Publicidad

Tabla de contenido
loading

Tabla de contenido