6.4 Fuse replacement
Prior to fuse replacement, ensure that multimeter is disconnected from
external voltage supply and the other connected instruments (such as
UUT, control instruments, etc.).
Exclusively use fuses of voltage and current values in compliance with
the technical data section.
Using auxiliary fuses, in particular short-circuiting fuse holders is
prohibited and can cause instrument destruction or serious bodily in-
jury of operator.
Switch of the instrument.
Loosen the screws on the instrument rear.
Lift the case cover.
Remove the defect fuse.
Insert new fuse.
Replace the case cover and retighten the screws
Fuse (mA) :
UNITEST Order. No.:
Fuse (A) :
UNITEST Order. No.:
F200mA 250V Ceramic 5x20mm
ES200MA00001
F10A 250V Ceramic 5x20mm
ES10000MAF01
43