Operational Conditions: Pump
Temperature +53°F (+12°C) to +93°F (+34°C)
Relative Humidity 20% to 70%
Atmospheric Pressure 0.82 atm (83.1kPa) to 1.05atm
(106.4 kPa)
Transport and Storage Conditions: Pump
Temperature -10°F (-23°C) to +130°F (+54°C)
Relative Humidity 20% to 90%
Atmospheric Pressure 0.82 atm (83.1 kPa) to 1.05atm
(106.4 kPa)
MEDICAL ELECTRICAL EQUIPMENT needs special precautions
regarding EMC and needs to be installed and put into service
according to the EMC provided in these Instructions for use.
Portable and mobile RF communications equipment can
affect MEDICAL ELECTRICAL EQUIPMENT.
Warning: The use of ACCESSORIES, transducers and
cables other than those specified, with the exception of
transducers and cabled sold by the MANUFACTURER of the
The Ameda Platinum Breast Pump is intended for use in
the electromagnetic environment specified below.
Emissions test
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions IEC 61000-3-2
Voltage fluctuations/flicker emissions
IEC 61000-3-3
The Ameda Platinum Breast Pump is intended for use in
the electromagnetic environment specified below.
I
test
mmunity
Electrostatic discharge
(ESD) IEC 61000-4-2
Electrical fast
transient/burst IEC
61000-4-4
Surge IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines IEC 61000-4-11
Power frequency
(50/60 Hz) magnetic
field IEC 61000-4-8
NOTE: U
is the a.c. mains voltage prior to application of the test level.
t
8
E N G L I S H
Guidance and manufacturer's declaration – electromagnetic emissions
Guidance and manufacturer's declaration – electromagnetic immunity
IEC 60601 test level
Compliance level
± 6 kV contact
± 6 kV contact
± 8 kV air
± 8 kV air
± 2 kV for power
± 2 kV for power
supply lines/± 1 kV for
supply lines/± 1 kV for
input/output lines
input/output lines
± 1 kV line(s) to line(s)
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
± 2 kV line(s) to earth
<5% U
(>95% dip in
<5% U
t
U
) for 0,5 cycle
U
t
40% U
(60% dip in
40% U
t
U
) for 5 cycles
U
t
70% U
(30% dip in
70% U
t
U
) for 25 cycles
U
t
<5% U
(>95% dip in
<5% U
t
U
) for 5 s
U
t
3 A/m
3 A/m
Weight: Approximately 9.75 lbs. (approximately 4.42 kg)
Dimensions: Length 9.75 in. (24.7 cm)/Width 10.5 in.
(26.7 cm)/Height 8.0 in. (20.3 cm)
Standards:
E325189 MEDICAL EQUIPMENT
WITH RESPECT TO ELECTRIC SHOCK,
FIRE AND MECHANICAL HAZARDS ONLY
IN ACCORDANCE WITH
IEC60601-1: 1988 + A1:1991 + A2:1995,
IEC60601-1-2, CAN/CSA C22.2 No. 601.1-M90
Ameda Platinum Breast Pump as replacement parts for
internal components, may result in increased EMISSIONS or
decreased IMMUNITY of the Ameda Platinum Breast Pump.
Warning: The Ameda Platinum Breast Pump should not
be used adjacent to or stacked with other equipment. If
adjacent or stacked use is necessary, the Ameda Platinum
Breast Pump should be observed to verify normal operation
in the configuration in which it will be used.
The customer or the user of the Ameda Platinum Breast
Pump should assure that it is used in such an environment.
Compliance
Electromagnetic environment – guidance
Group 1
The Ameda Platinum Breast Pump uses RF energy
only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
Class B
The Ameda Platinum Breast Pump is suitable
for use in all establishments, including domestic
Class A
establishments and those directly connected to
Complies
the public low-voltage power supply network that
supplies buildings used for domestic purposes.
The customer or the user of the Ameda Platinum Breast
Pump should assure that it is used in such an environment.
Electromagnetic environment – guidance
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30%.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
(>95% dip in
Mains power quality should be that of a typical
t
) for 0,5 cycle
commercial or hospital environment. If the user
t
(60% dip in
of the Ameda Platinum Breast Pump requires
t
) for 5 cycles
continued operation during power mains
t
(30% dip in
interruptions, it is recommended that the Ameda
t
) for 25 cycles
Platinum Breast Pump be powered from an
t
(>95% dip in
uninterruptible power supply.
t
) for 5 s
t
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.