Set The Sample Probe Position - Hach HIAC 9703+ Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 39
Figure 7 Attach the syringe
1 Plunger bolt
2 Syringe
1. Push the ON/OFF switch to turn the instrument on.
2. Push the LOAD button under the syringe compartment to set the
plunger bolt at the lowest point. Refer to
3. Loosen and remove the plunger bolt from the syringe drive.
4. Hold the syringe and push the bolt through the plunger hole.
5. Push the bolt in the syringe drive hole and turn the bolt by hand to
tighten.
6. Pull the syringe barrel up and install the syringe threads in the valve.
Figure
7.
3 Syringe drive
4 Valve
7. Turn the syringe by hand to tighten.
Note: Do not overtighten.

Set the sample probe position

The lift mechanism for the sample probe can be set to automatically go
to the correct position for sampling.
English 17