2.4 Recycling/Disposal
By disposing of this product and its packaging in the correct manner you are helping to
prevent potential negative impacts on people and the environment that could be caused
by incorrect disposal. Comply with local regulations and bring the units to be disposed of
to a proper recycling facility.
3 Operating and Safety Information
3.1 Qualified Personnel
Installation and maintenance of the filter system may be carried out only by trained or
authorised personnel.
3.2 Correct Use
The product can only be operated properly and safely if it is installed, used and serviced
in the manner described in this Manual. Use only filter cartridges suitable for this appliance.
the HYDROS Treatment System and system-specific filter cartridges may only
Note:
be used with the C61wd G4 and M2 G4 units.
3.3 Liability Exclusion
Installation must be carried out in accordance with the instructions in this Manual. SciCan
shall not be held liable for any damage, including subsequent damage, arising from the
incorrect installation or use of the product.
3.4 Specific Safety Information
Only drinking water within the intake water temperature range listed in Chapter 10
may be used as intake water for the HYDROS water filter system.
No microbiologically impaired water or water of unknown microbiological quality may
be used without appropriate sterilisation.
SciCan recommends that the filter system shall not be decommissioned for a long
period. After stagnation times of over 4 weeks, the tank should be flushed or a
pre-rinse executed. Please also note the maximum usage period of the filter cartridge
is 12 months (Chapter 6).
The filter system is not resistant to heavily concentrated cleaning agents (e.g. bleach,
chlorinated solvents, heavy oxidants) and must not come into contact with them.
The filter system must not be opened or dismantled during operation.
The filter cartridge must not be opened.
The pressure vessel and the pressure vessel lid of the filter systems have a service
life of up to ten years (from the date of installation), provided that they are installed
and used correctly and the operating conditions outlined in the Technical Data chapter
are adhered to. They must always be replaced after a maximum of ten years. The
hoses must be replaced in rotation after a maximum of five years.