Dantec Keypoint
Call qualified service personnel to conduct
at least a functional test and additionally a
safety check including 1) an insulation test;
2) a ground continuity test; and 3) a leakage
current test according to IEC 60601-1.
The protection is likely to be impaired if,
for example, the device:
- Shows visible damage;
- Fails to perform the intended function(s);
- Has been subjected to severe transport
stress.
Connection Panel
Models: 9031D040x / 9031D041x
CAUTION In the USA: If 240V~, only center-
tapped, single phase circuit.
NOTE Unplugging the power line cable from
the mains input on the Isolating Transformer,
disconnects the mains power to all supplied
devices.
(1)
Mains Input
Plug the shielded mains power cord sup-
plied with the system into this socket.
NOTE Only use a shielded power cord from
Alpine Biomed.
Rating Frequency: 50/60Hz.
-Continuous Operation: max. 1000VA
-Momentary Operation: max 1500VA
9031D040x: 100/120V~
9031D041x: 200/240V~
(2)
Fuse
F1 and F2 Mains
9031D040x: 110V/120V: T10A / 250V~
9031D041x: 200/240V: T6.3A / 250V~
(3)
M4 Thread for Electrical Enclosure Con-
tact and Additional Protective Earth.
Potential equalization
(4)
Where potential equalization is necessary or
desirable, a potential equalization conductor
may be connected to this connector as per
DIN 42 801.
(5)
Spring Retaining Clamp for Power Cord
Place the spring retaining clamp around the
power cord to ensure a safe connection.
(6)
Auxiliary Outlet-6 Outlets: 50/60Hz
9031D040x:
100-120V~ total max.
-Continuous Operation 7.4A
-Momentary Operation 12.4A
7.4A equals 850VA, and 12.4A equals
1425VA at 115V~
9031D041x:
200-240V~ total max.
-Continuous Operation: 3.7A
-Momentary Operation: 6.2A
3.7A equals 850VA and 6.2A equals
1425VA at 230V~
(7)
SN: Device Serial Number
(8)
REF: Device Reference Number
Environmental Conditions
Operating Environment
Temperature:
+10°C to +35°C (+50°F to +95°F)
Relative Humidity:
20% to 80% (non-condensing)
Altitude:
-15m to 3,000m
Pressure:
701hPa to 1060 hPa
Storage/Transport Environment
Temperature:
-40°C to +65°C (-220°F to 149°F)
Relative Humidity:
20% to 80% (non-condensing)
Altitude:
-15m to 10,600m
Pressure:
241hPa to 1060 hPa
Maintenance
The maintenance that can be performed by the
operator is limited to cleaning and disinfecting
the device. Any maintenance inside the device
must be performed by qualified service
personnel only.
Before cleaning the device(s), switch off at
the mains.
Clean the device with a clean, slightly
damp cloth, and wipe it dry.
CAUTION Do not use solvent silicon-based or
abrasive cleaning agents.
CAUTION Make sure that no liquids enter the
device through the connectors and other
openings in the enclosure.
– 6 –