Bien Air OPTIMA Guia Del Usuario página 39

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 42
OPTIMA INT
List of errors & troubleshooting
Communication timeout.
Message
Error(s): #3 comm. timeout
Root cause
Communication failed with the DMCX/DMX2/DMX3 board.
Action
Check DMCX/DMX2/DMX3 connection (red 8-pin connector).
Contact retailer.
DMCX/DMX2/DMX3 board over-temperature.
Message
Error(s): #4 temperature
Root cause
DMCX/DMX2/DMX3 board too hot (electronic device overheating).
Action
Wait until the DMCX/DMX2/DMX3 cools down.
Contact retailer.
DMCX/DMX2/DMX3 board under-voltage.
Message
Error(s): #5 under voltage
Root cause
DMCX/DMX2/DMX3 board 32V power supply detected voltage too low.
Action
Contact retailer.
DMCX/DMX2/DMX3 board over-voltage.
Message
Error(s): #6 over voltage
Root cause
DMCX/DMX2/DMX3 board 32V power supply detected voltage too high.
Action
Contact retailer.
DMCX/DMX2/DMX3 unknown failure.
Message
Error(s): #0 unknown
Root cause
DMCX/DMX2/DMX3 unknown failure (unidentified cause).
Action
Contact Bien-Air Dental SA.
37