450 Series Waterproof Handheld Meter
Dissolved Oxygen Operation Instructions
Models:
DO 450
DO/Temperature
pH/mV/DO/Temperature
Getting Started/Connections
After installing (2) AA batteries and/or connecting the optional
110/220 VAC power supply, connect the desired sensors to the
corresponding ports.
DO: 9-Pin RDO/ATC
ATC: 6-Pin ATC
(PD 450 only)
A/C Power
Adapter
RS232 – PC or Printer
The stand can be extended as shown above or used for
wall-mounting.
Keypad Functions
Press once to power ON in the mode that was
previously used. Press again to turn backlight
on for one minute or off. Hold for 3 seconds to
power OFF.
Toggle between measurement and calibration
modes. In SETUP mode, BACK serves to return to
the previous menu option or setting.
Confirm calibration values in CAL mode.
Confirm selections in SETUP mode.
Freeze or release the measured reading.
Customize instrument settings and preferences.
(See also Setup Programs)
Toggle between available measurement types.
Save measurement into memory. Increase value
or scroll up in SETUP or manual calibration.
Recall saved values from memory. Decrease value
or scroll down in SETUP or manual calibration.
Send output data to printer or computer.
Setup Programs
To access the settings below, press SETUP. Up/down arrows
will display the available options. Press ENTER to accept the
desired setting, or BACK to return to the previous option
and/or exit.
Configuration Options
• Ready indicator ON / OFF / or Automatic HOLD when stable
• Choose ºCelsius or ºFahrenheit
RDO Setup Options
• Choose units (% saturation, mg/L, or ppm)
• Choose number of calibration points 1 (no zero) or 2
(includes zero)
• Set compensation pressure in kPa (60.0 - 113.3), or mmHg,
(450.0 - 850.0), salinity correction (0.0 - 45.0 ppt)
RDO Info
• Remaining Cap Life (# Days)
• Probe Serial Number (####)
• Year of Probe Manufacture (####)
• Date of Probe Manufacture (DD.MM)
Select Calibration Due Reminder
PD 450
• Set number of days from 0-60
View Calibration Data (for the parameter being measured)
• Press ENTER to view each point that is calibrated.
View Electrode Data (for the parameter being measured)
• Press ENTER to view electrode data (Torr / % saturation).
System Settings
• Data Logging:
MANUAL upon key press only
TIMED interval. Choose (SEC / MIN / HOUR) interval.
• Automatic shut off after 10 minutes. Choose ON or OFF.
• Clock Settings:
Date: Choose USA (MM/DD/YYYY) or Euro (DD/MM/YYYY).
Time: Choose (24HR or 12HR). If 12HR, choose AM or PM.
BNC: pH, mV/ORP
or Ion Selective
• Set Printer Type:
(PD 450 only)
CSV (Comma Separated Values) – best format for computer.
Printer (Text) – best format for printer.
Choose Manual (MAN) upon key press or TIMED interval.
If timed, choose (SEC / MIN / HOUR).
Reset
Output (USB or
• NO. Exits from reset menu options without action.
• FACTORY RESET. Returns all settings except date/time
Communication)
and ATC calibration to factory default values after ENTER is
pressed then restarts meter.
• DATA RESET. Erases data stored in memory while retaining
other settings after ENTER is pressed.
• CALIBRATION RESET. Erases non-ATC calibration data while
retaining other settings after ENTER is pressed.
Dissolved Oxygen Calibration
Perform daily calibration for best results. Dissolved oxygen
levels vary with temperature, barometric pressure, and salinity,
so calibration must be performed with consideration of these
factors. It is necessary to adjust the barometric pressure
and salinity values and calibrate the temperature prior to
performing any DO calibration or measurement. For probe
operation, refer to RDO Operation Instructions.
The 450 series can accept one or two calibration values;
typically 100 % using saturated air or air-saturated water, and
0 % using zero oxygen solution. If calibrating for 0 % oxygen,
note that the meter will take several minutes to reach 0 %
saturation value and constant stirring is not required.
100 % Saturation Automatic Calibration
1. Rinse the probe well with clean water. From % saturation
mode, press CAL.
2. The primary display will show the calibration value, while
the secondary display shows the unadjusted value and the
lower display shows the measured temperature.
3. Place the probe in the optional calibration chamber or a
saturated oxygen environment (air or air saturated water) &
wait for the reading to stabilize. Press ENTER to accept.
0 % Saturation Automatic Calibration
1. Rinse the probe well with clean water. From % saturation
mode, press CAL.
2. The primary display will show the calibration value, while
the secondary display shows the unadjusted value and the
lower display shows the measured temperature.
3. Submerse probe in zero oxygen calibration solution and
stir. Wait for the reading to stabilize – this can often take
more than 10 minutes. Wait for the reading to stabilize.
Press ENTER to accept.
Part of Thermo Fisher Scientific
For Sales & Service Contact
2650 E. 40th Ave. • Denver, CO 80205
Phone 303-320-4764 • Fax 303-322-7242
1-800-833-7958
www.geotechenv.com
mg/L or ppm Manual Calibration
1. Rinse the probe well with clean water. From mg/L or ppm
mode, press CAL.
2. The primary display will show the calibration value, while
the secondary display shows the unadjusted value and the
lower display shows the measured temperature.
3. Place the probe into a sample of known oxygen
concentration (i.e. determined by titration or another
instrument). Wait for the reading to stabilize. Press t/s to
adjust the value and ENTER to accept.
Temperature Calibration/Manual ATC
1. Press CAL from any measurement, then press MODE.
2. Skip to step 3 for manual ATC, otherwise, dip the temperature
sensor into a solution with a known accurate temperature.
The upper display shows the active temperature while
the lower display shows the factory default temperature
without adjustment.
3. Use up/down arrows to adjust the upper display. Press
ENTER to accept the calibration temperature. The
maximum adjustable value is ±10 ºC (or ±18 ºF) from
factory default.
Error Messages
"ERR" will appear when an error condition exists or the
incorrect key is pressed. Common examples include:
• Pressing ENTER during calibration before the "READY"
indicator appears. Wait for the "READY" indicator before
pressing ENTER.
• UR (Under Range) • OR (Over Range)
Intended Use, Maintenance & Precautions
These handheld meters use sensors to detect various parameters
for water-based measurements. For routine maintenance
disconnect the power cord or battery, then dust or wipe the
display using a damp cloth. If necessary, warm water or a mild
water based detergent can be used. Immediately remove any
spilled substance from contact with the meter using the proper
cleaning procedure for the type of spill.
• Do not use this equipment in potentially explosive atmospheres.
• Refer to the electrode instructions for use, storage and cleaning.
• Ensure that no liquid enters the instrument.
• Do not use any aggressive cleaning chemicals (solvents or
similar agents).
• There are no user serviceable parts inside. Attempts to
service internal parts may void the warranty.
• WARNING: No modification of this equipment is allowed.
Instrument Operating Conditions
Operating Ambient Temp.
5 to 45 ºC
Operating Relative Humidity 5 to 85 %, non-condensing
Storage Temp.
-20 to +60 ºC
Storage Relative Humidity
5 to 85 %, non-condensing
Pollution
Degree 2
Overvoltage
Category II
Weight
500 g
Size (L x W x H)
21.15 x 9.87 x 5.85 cm
Regulatory & Safety
CE, TUV 3-1, FCC Class A
Power Rating
DC Input: 9 VDC 1 A
2 x AA (LR6) 1.5 V batteries
Battery Requirement
(replace batteries when
battery sign blinks)
Vibration
Shipping/handling per ISTA #1A
Drop test in packaging per
Shock
ISTA #1A
Enclosure (Designed To Meet) IP67 (using rubber covers)
Universal Power Adapter Operating Conditions
Operating Ambient Temp.
0 to 50 ºC
Operating Relative Humidity 0 to 90 %, non-condensing
Storage Temp.
-20 to +75 ºC
Storage Relative Humidity
0 to 90 %, non-condensing
Pollution
Degree 2
Overvoltage
Category II
I/P: 100 - 240 V, 50/60 Hz, 0.3A
Power Rating
O/P: 9 VDC 1 A
Eutech Instruments Pte Ltd
Oakton Instruments
Blk 55, Ayer Rajah Crescent,
625 East Bunker Court,
#04-16/24, Singapore 139949
Vernon Hills, IL, 60061, USA
Tel: (65) 6778-6876
Tel: 1-888-462-5866
Fax: (65) 6778-0086
Fax: 1-847-247-2984
eutech@thermofisher.com
info@4oakton.com
www.eutechinst.com
www.4oakton.com
68X651903 08/14 Rev0

Publicidad

loading

Resumen de contenidos para Oakton 450 Serie

  • Página 1 3. Submerse probe in zero oxygen calibration solution and and/or exit. stir. Wait for the reading to stabilize – this can often take Eutech Instruments Pte Ltd Oakton Instruments Configuration Options more than 10 minutes. Wait for the reading to stabilize. Blk 55, Ayer Rajah Crescent, 625 East Bunker Court, •...
  • Página 2 2. La visualización primaria va a mostrar el valor de calibración mientras estable Eutech Instruments Pte Ltd Oakton Instruments que la visualización secundaria muestra el valor no ajustado y la • Elija ºCelsius o ºFahrenheit visualización inferior muestra la temperatura medida.
  • Página 3 Attendre que la lecture se stabilise; ceci peut souvent prendre plus de 10 minutes. Attendre que la lecture se stabilise. Appuyer sur ENTER • Indicateur de préparation ON / OFF / ou automatiquement figé (HOLD) Eutech Instruments Pte Ltd Oakton Instruments (Entrée) pour valider. si stable Blk 55, Ayer Rajah Crescent, 625 East Bunker Court, •...
  • Página 4 21,15 x 9,87 x 5,85 cm RDO-Bedienungsanleitung. Den gemessenen Wert einfrieren oder freigeben. Behördliche Vorschriften und Die 450 Serie kann einen oder zwei Kalibrierungswerte akzeptieren; CE, TUV 3-1, FCC Klasse A Sicherheit Instrumenteneinstellungen und Präferenzen anpassen. normalerweise 100 % unter Verwendung von gesättigter Luft oder...