Control with Band Limitation

10 Monitoring of Measuring Circuit

Lower Display
Range limit
Meas. range
3.3 mV/V
3.0 mV/V
2.0 mV/V
1.5 mV/V
1.0 mV/V
Indication
blinking
( = is detected)
1) Wire short circuit between Signal input + and Signal input – is not detected.
2) Wire break with Sensor supply + or Sensor supply – is not detected.
Upper Display
Lower Meas.
Range limit
Range limit
blinking
blinking
Upper Meas.
Probe or wire
Range limit
short-circuit
Probe or wire
break
1) 2)
blinking
blinking