Recommended separation distances between portable and mobile RF
communications equipment and the Power Neb Ultra Compressor Nebulizer 18090
The 18090 is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the 18090 can help prevent
electromagnetic interference by maintaining a minimum distance between portable and
mobile RF communications equipment (transmitters) and the 18090 as recommended
below, according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter
W
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in metres (m) can be estimated using the equation applicable to the frequency of the trans-
mitter, where P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
14
.............................................................................................................
DRIVE COMPRESSOR/NEBULIZER MODEL 18090-BE / 18091-BE
............................................................................................................
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d = 1.2 √ P
0.12
0.38
1.20
3.79
12.00
m
80 MHz to 800 MHz
d = 1.2 √ P
0.12
0.38
1.20
3.79
12.00
800 MHz to 2,5 GHz
d = 2.3 √ P
0.23
0.73
2.30
7.27
23.00