6. Operating Conditions
Operating Temperature:
Relative humidity:
7. Storage and Transport Conditions
Storage temperature
Relative humidity
Storage conditions
8. Frequently Asked Questions
Question:
What happens if I connect the mask-end cuff to the humidifier and the humidifier-
end cuff to the mask?
Answer:
The Hybernite System works exactly as if the tube was connected correctly (mask
cuff to mask and humidifier cuff to humidifier). However, for safety reasons and to
ensure maximum comfort, connect the tube as described in the user's manual.
Question:
Can I use the tube without a power supply?
Answer:
Yes, the tube will then works like a standard tube for respiratory therapy devices.
Respiratory air condensation can however form in the tube.
Question:
When do I have to replace the tube?
Answer:
The tube must not be used for longer than 12 months. If the tube is damaged or a
leak or malfunction occurs, then the tube should be replaced immediately.
Question:
Which humidifier or flow generator can I use the Hybernite System with?
Answer:
The cuffs of the Hybernite have the same diameter as standard tubes for respiratory
therapy systems. Therefore, you can use the tube with all humidifiers that are
intended for the connection of a standard tube: examples: ResMed HumidAire 2i;
ResMed HumidAire 3i; ResMed H4i; ResMed H5i; Philips Respironics System One
humidifier; Respironics M-series humidifier; Respironics Legacy Series humidifiers;
Breas HA20 Humidifier; Devilbiss Heated Humidification systems; Weinmann
SOMNOcomfort 2; Weinmann SOMNOsoft 2; Weinmann SOMNObalance;
Hoffrichter Point2; Sefam Dreamstar; Sefam Sandman; Apex; BMC; etc
Question:
How do I get maximum comfort from using the Hybernite Rainout Control System?
Answer:
The maximum comfort will be different for every patient. The patent is advised to
change the settings of the humidifier to find the zone of optimal comfort. It can
be advised to increase the settings of the humidifier to the mid or top third of the
range (e.g. setting 3, 4 or 5 if the range is from 1 to 5 or settings 4, 5 or 6 if the
range is from 1 to 6).
9. Product Specifications
Hybernite HBT 300 Heated Breathing Tubing
Length
Internal Diameter
Flow resistance
Tubing Compliance Standards
Water resistance
Electrical Safety Standard
5°C – 30°C (41°F – 86°F)
5% – 95%
- 20 °C (-4°F) ... + 60 °C (120°F)
5 % ... 95 %
The Hybernite System should be transported and stored in the original
packaging and protected from dust and exposure to direct sunlight.
1.83 m
19 mm
60 l∙min
-1
ISO 5367:2000
ISO 8185:2007
IPx3
EN IEC 60601-1
Type B Applied Part ( )
≤ 0.1 kPa