• Description of alarms
Alarm
code
0
Microcontroller memory alarm
32
Microcontroller memory alarm
33
Short-circuit in one of the two motors
34
Short-circuit before the motors' starting
35
Condensers have not been charged
36
Overtemperature (temperature has exceeded
the max. allowed limit)
37
Blower overcurrent (the blower has exceeded
the current limit)
39
Condensers overvoltage (the condensers have
exceeded the max.voltage limit)
40
Centrifuge short-circuit
41
PC-Board short-circuit on the centrifuge output
42
Centrifuge instant overcurrent (the centrifuge
has exceeded the current limit)
43
Centrifuge time delayed overcurrent
(the centrifuge has exceeded the current limit)
46
Amalgam level > 95%
47
Amalgam level > 100%
48
The vacuum detecting tube is disconnected
IMPORTANT NOTICES
• Appliances are guaranteed for one year from the date of sale, provided that the warranty slip is
returned to the manufacturer with date of sale, retailer's and customer's name.
• Warranty and manufacturer's liability cease in case appliances are treated with products which are
unsuitable or different from those recommended by the manufacturer and also in case appliances
are improperly used or tampered with operations of any kind carried out by people who are not
authorized by the manufacturer.
• The manufacturer, concessionaires, agents and authorized technicians are at customers' disposal
for advice and assistance and to supply literature, spare parts and anything useful.
• The manufacturer reserves the right to modify the products for improvements, for technical,
normative and functional reasons or for problems due to the availability of products or semi-fi nished
products, without prior notice.
• Our updated manuals are available at www.cattani.it.
We recommend they are consulted especially for updates concerning safety.
38
DESCRIPTION
SOLUTION
Contact the technician
Contact the technician
Identify where the short-circuit is coming from
and eliminate it
Contact the technician (the PC-Board is likely to
be damaged)
Contact the technician (replace the PC-Board)
Ventilate the plant room
Check operation of the cooling fan
Check the operation of the blower (seized or
excessive friction)
Check the mains voltage (max. 260V)
Eliminate the short-circuit
Replace the PC-Board
Eliminate siphons in the piping or check the
centrifuge operation (seized or excessive friction)
Eliminate siphons in the piping or check the
centrifuge operation (seized or excessive friction)
Replace the amalgam container as soon as
possible
Replace the amalgam container
Connect the vacuum detecting tube to the
centrifuge