Dwyer Magnehelic 2000-0 3% Instrucciones Y Lista De Partes

Manometro diferencial

Publicidad

Idiomas disponibles

Idiomas disponibles

CIRCLE FOR PANEL MOUNTING
11/16
[17.46]
DWYER INSTRUMENTS, INC.
P.O. BOX 373 • MICHIGAN CITY, INDIANA 46361 U.S.A.
Magnehelic
®
O P E R A T I N G I N S T R U C T I O N S
Ø4-1/8 [104.78] BOLT
120° APART
1/8 NPT HIGH
PRESSURE
1-1/4
1-3/4
[31.75]
[44.45]
1/8 NPT LOW
PRESSURE
7/16
[11.11]
1-11/16
[42.86]
B U L L E T I N N O .
Differential Pressure Gage
SPECIFICATIONS
Dimensions: 4-3/4 dia. x 2-3/16 deep.
Weight: 1 lb. 2 oz.
Finished: Baked dark gray enamel.
Connections: 1/8 NPT high and low pressure
taps, duplicated, one pair side and one
pair back.
Accuracy: Plus or minus 2% of full scale, at
70°F. (Model 2000-0, 3%; 2000-00, 4%).
Pressure Rating: 15 PSI (0,35 bar)
Ambient Temperature Range: 20° to 140°F
(-7 to 60°C).
Standard gage accessories include two 1/8
NPT plugs for duplicate pressure taps,
two 1/8 NPT pipe thread to rubber tubing
adapters, and three flush mounting
adapters with screws.
Caution: For use with air or compatible
gases only.
For repeated over-ranging or high cycle rates,
contact factory.
Not for use with Hydrogen gas. Dangerous
reactions will occur.
1/2
[12.70]
1/8 NPT HIGH
PRESSURE
1/8 NPT LOW
PRESSURE
1/2
[12.70]
Phone: 219/879-8000
Fax: 219/872-9057
Lit-by-Fax: 888/891-4963
A-27
Ø4-3/4
[120.65]
www.dwyer-inst.com
e-mail: info@dwyer-inst.com

Publicidad

Tabla de contenido
loading

Resumen de contenidos para Dwyer Magnehelic 2000-0 3%

  • Página 1 1/8 NPT LOW 1-1/4 PRESSURE 1-3/4 [31.75] [44.45] 1/8 NPT LOW PRESSURE 7/16 11/16 Ø4-3/4 [11.11] [17.46] [120.65] 1-11/16 [42.86] [12.70] Phone: 219/879-8000 www.dwyer-inst.com DWYER INSTRUMENTS, INC. Fax: 219/872-9057 e-mail: info@dwyer-inst.com P.O. BOX 373 • MICHIGAN CITY, INDIANA 46361 U.S.A. Lit-by-Fax: 888/891-4963...
  • Página 2 ® MAGNEHELIC INSTALLATION 5. To zero the gage after Select a location free from excessive vi- installation bration and where the ambient temperature Set the indicating pointer exactly on the will not exceed 140°F. Also, avoid direct zero mark, using the external zero adjust sunlight which accelerates discoloration of screw on the cover at the bottom.
  • Página 3 Consult factory for assistance on unusual applications or conditions. Use with air or compatible gases only. ©Copyright 2001 Dwyer Instruments, Inc. Printed in U.S.A. 4/01 FR# 12-440212-00 Rev. 3 Phone: 219/879-8000 www.dwyer-inst.com DWYER INSTRUMENTS, INC.
  • Página 4 Consult factory for assistance on unusual applications or conditions. Use with air or compatible gases only. ©Copyright 2001 Dwyer Instruments, Inc. Printed in U.S.A. 4/01 FR# 12-440212-00 Rev. 3 Phone: 219/879-8000 www.dwyer-inst.com DWYER INSTRUMENTS, INC.
  • Página 5: Especificaciones

    1/8 NPT LOW 1-1/4 PRESSURE 1-3/4 [31.75] [44.45] 1/8 NPT LOW PRESSURE 7/16 11/16 Ø4-3/4 [11.11] [17.46] [120.65] 1-11/16 [42.86] [12.70] Phone: 219/879-8000 www.dwyer-inst.com DWYER INSTRUMENTS, INC. Fax: 219/872-9057 e-mail: info@dwyer-inst.com P.O. BOX 373 • MICHIGAN CITY, INDIANA 46361 U.S.A. Lit-by-Fax: 888/891-4963...
  • Página 6: Instalación

    INSTALACIÓN 5. Puesta a Cero Después de Seleccione un lugar libe de exceso de vi- Instalar braciones, y donde la temperatura ambiente Deje las conexiones de presión abiertas a no supere los 60°C. Evite luz solar directa, atmósfera y ajuste a cero desde tornillo del para evitar decoloración de la cubierta plás- panel frontal.
  • Página 7: Mantenimiento

    En caso necesario envie el instru- Verificación de Calibración: Desconecte el mento con transporte pago a: instrumento de proceso, ventee a atmósfera Dwyer Instruments, Inc. y deje escurrir condensados. Utilice un Attn: Repair Department manómetro de calidad y exactitud conoci- 102 Indiana Highway 212 das, y de rango adecuado.

Este manual también es adecuado para:

Magnehelic 2000-00 4%