ELECTROMAGNETIC COMPATIBILITY
PRECAUTIONS:
Medical
electrical
{
special precautions regarding electromagnetic
compatibility (EMC) and needs to be installed
on this cart according to the EMC information
provided below and in the respective equipment
instructions.
The essential performance of the cart is to
{
maintain continued operation, providing power
to support equipment.
The combination of cart and equipment is
{
intended for use by healthcare professionals only.
The system may cause radio interference or may
disrupt the operation of nearby equipment. It may
be necessary to take mitigation measures, such as
reorienting or relocating the cart or shielding the
location.
Guidance and Manufacturer's Declaration: Electromagnetic Emissions
Video Carts are intended for use in the electromagnetic environment specified below. The customer or
of the user of the Video Cart should ensure that it is used in such an environment.
Emissions Test
Compliance
RF emissions
CISPR11
RF emissions
CISPR11
Harmonic emissions
Not applicable
IEC61000-3-2
Voltage fluctuations/
Not applicable
flicker emissions
IEC61000-3-3
2017/12
equipment
Electromagnetic Environment -
Guidance
Group 1
The video cart uses RF energy only for its
internal function. Therefore, its RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment.
Based on the product description there are
no electrical analog or digital circuitry that
could generate or conductively couple RF
energy. Therefore conducted and radiated
emissions testing was not performed.
Class A
Based on the product description, there are
no electrical analog or digital circuitry that
could generate harmonic distortion on to
the AC power mains, generate flicker on the
AC power mains, or be affected by voltage.
Therefore the harmonic distortion testing
was not performed.
The video cart is suitable for use in all
establishments other than domestic and
those directly connected to the public low-
voltage power supply network that supplies
buildings used for domestic purposes.
needs
Using accessories, transducers and cables
{
other than those specified for use with the cart
and the equipment loaded on the cart may
result in increased emissions or decreased
immunity to such emissions.
When stacking devices on the cart, observe
{
the equipment to verify normal operation in
the configuration in which it will be used.
This system was not tested for immunity
{
to electromagnetic disturbances.
Portable and mobile RF communications
{
equipment can affect the performance of
medical electrical equipment loaded on this
cart.
Guidance and Manufacturer's Declaration: Electromagnetic Immunity
The Video Cart is intended for use in the electromagnetic environment specified below.
The customer or the user of the Video Cart should ensure that it is used in such an environment.
Immunity Test
Electrostatic
Discharge (ESD)
IEC61000-4-2
Electrical fast
transient/burst
IEC61000-4-4
Surge
IEC61000-4-5
Voltage
dips, short
interruptions
and voltage
variations on
power supply
input lines
IEC61000-4-11
Power frequency
(50/60Hz)
magnetic eld
IEC 61000-4-8
NOTE: Ut is the a.c. mains voltage prior to application of the test level.
2
IEC 60601Test
Compliance Level
Level
±6kV contact
not applicable
±8kV air
±2kV for power
±2kVAC Mains
supply lines
±1kV I/O Lines
±1kV for input/
5/50
output lines
5kHz
±1kV differential
±1 kV
mode
Line to Line
±2kV common mode
±2 kV
Line to Ground
<5% Ut (>95% dip in
not applicable
Ut) for 0.5 cycle 40%
Ut (60% dip in Ut)
for 5 cycles 70% Ut
(30% dip in Ut) for
25 cycles <5% Ut
(>95% dip in Ut) for
5 sec.
3 A/m
N/A
Electromagnetic
Environment: Guidance
Based on the product
description there are no
electrical analog or digital
circuitry that could be affected
by an ESD directly or indirectly.
Therefore ESD testing was not
performed
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment.
Based on the product
description there are no
electrical analog or digital
circuitry that could be
affected by voltage dips, short
interruptions and voltage
variations on the AC power
mains. Therefore no testing was
performed.
Based on the product
description there are no
electrical analog or digital
circuitry that could be affected
by magnetic fields. Therefore no
testing was performed.
1000-400-701-EN REV. R