Power On/Off; Setup Mode; Measurement Mode - RS Pro RS-96 Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

Mini Particle Counter / English
4. Specifications
Particle Counter
Channel
2.5μm, 10um
PM2.5: 0~2000ug/m
Mass Concentration
PM10: 0~2000ug/m
3
Resolution
1ug/m
Air Temperature Measure
Air Temp Range
0~50°C/ 32~122°F
Resolution
0.1°C/°F
Basic Accuracy
±1°C/2°F
Humidity Range
0 to 100%RH
Resolution
0.1%RH
±5%RH 0~20%RH, 80~100%RH
Basic Accuracy
±3.5%RH 20~80%RH
5. Power On/ Off
On the power off mode, press and hold
button until the LCD is on, then the unit will
power on.
On the power on mode, press and hold
button until the LCD is off, then the unit will
power off.

6. Measurement Mode

On the power on mode, you can press the button to start
PM10, the top left corner of the LCD display "Counting", the top right corner of the LCD
display count down, LCD main display PM2.5 and PM10 data and temperature & humidity
readings are at the bottom of LCD.
2
3
3
measuring PM2.5 and
19/12/2017 Version No. 001
Press the button again to stop the measurement. The top left corner of the LCD display
will show "Stopped". The LCD display will show the last measurement data. Data will be
automatically save to the instrument memory, which can store up to 5000 data.

7. Setup mode

Powering on the instrument. Long press button
as shown below:
Press the
button and the
button to select the menu option needed. Then
press button
to enter into the appropriate settings page.
19/12/2017 Version No. 001
Mini Particle Counter / English
to enter into the system setup mode
3

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

155-8903