INSTRUCTION MANUAL CODE 80002 REV E
"SENSOTERM II"
J
.P. SELECTA s.a.u.
STARTING UP :
AGIMATIC-S:
Once the indications in the Installation section have been made,
switch on the main switch on the rear part, place the recipient to
be stirred on the top plate, let the teflon-coated magnetic stirring
bar drop into the recipient, turn on the "MOTOR" switch and adjust
speed with "RPM" knob.
AGIMATIC-N:
For its starting up operation and to select the speed, do as the
above model, AGIMATIC-S. There is no heating until motor is
working.
Then switch on the HEATER switch, and with the regulator for
heating power, select the required "%". The yellow lamp will show
the working of the heater.
AGIMATIC-E:
To start it working and to adjust speed, follow the above indications.
With the regulator "HEATER" the suitable working temperature may
be selected with the "ºC" knob.
LIQUID TEMPERATURE CONTROL
(N & E models)
1. Connection to work without external connection of a temperature
2. Connection for contact thermometer.
• Put the thermometer with the bulb submerged into the liquid.
• Set the temperature turning the superior part of the thermometer.
Figure 1
• The connection is as shown on figure 1.
3. Connection for regulator "SENSOTERM II" (1001009)
• Put the probe submerged into the liquid. Connect the probe in
• Set the temperature on the «SENSOTERM II».
• Set the power rate or temperature on the stirrer at maximum.
• The connection is as shown on figure 2.
Figure 2
Ctra. NII Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.es - website: www.jpselecta.es
control device.
Connect the thermometer in the stirrer rear part plug.
The stirrer must be configured to work at maximum of power
rate or temperature.
the stirrer rear part plug.
04/2020
Tel 34 93 770 08 77 Fax 34 93 770 23 62
Page: 15