INSTRUCTION MANUAL CODE 80189 REV D
J
.P. SELECTA s.a.u.
Jun/2018
10 Technical specifications
Mains voltage:
Electrical power:
Maximum consumption:
Fuses:
Power supply cord:
Weight (Net):
Dimensions: W x L x H
Water reservoir volume:
NaOH, H3BO3 and HCl tank volume:
Nitrogen recovery:
Detection range:
Reproducibility: relative standard deviation
Normality range
Detected nitrogen resolution:
Destillation rate:
Analysis time for 50mg of N:
Analysis time for 100mg of N:
Cooling water entry connector:
Cooling water output connector:
Remaining steam output connector:
Dosification for NaOH pump:
Dosification for H3BO3 pump:
Sound level:
Ambient conditions:
Overvoltage category:
Pollution degree:
Glass material:
Door and front material:
Carriage material:
Chamber material:
Steam generator material:
Hosepipe material:
RS-232 output:
Autovía A-2 Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.es - website: http://www.jpselecta.es
(It can be modified without notice)
230V 50/60Hz (See plate)
Other power supply voltages could be
available (ask to your dealer)
1900W (See plate)
8.5 A
10 A fast. dimensions: 20x5 mm
Phase, Neutral, Earth (section 1mm
35 Kg
500 x 490 x 740 mm
6 litres
2 litres
>99.5%
0.1 to 200 mg
< 1%
0.010 to 1.000 N
0.0001
approx. 30 / minute
6 minutes approx.
9 minutes approx.
3/4''
3/4''
3/4''
0 to 250 ml ±5%
50 to 65 ml ±5%
Lower than 60 dBA
Inner use.
Altitude up to 2000m
Temperature from 5ºC to 40ºC.
Relative maximum damp 80% for above
31ºC temperatures, dicreasing lineally to
50% for 40ºC.
Category II
2
Borosilicate 3.3
ABS injected
Iron with overlap
Aluminium with RILSAN coating chemical
resistant.
Stainless steel.
Silicone, synthetic fluorated elastomer.
Connector 9 pins male. Pin 2 and 3 inter-
changed, pin 5 common. Baud rate 2400,
no parity, 8 bit data, 1 bit stop.
Tel 34 937 700 877 Fax 34 937 702 362
Page: 60
)
2