– Do not damage the contact surfaces on the mPack UNPLUGGED
or the contact pins in the EN50 UNPLUGGED charger with pointed
or sharp-edged objects.
– Clean the contacts of EN 50 UNPLUGGED and mPack
UNPLUGGED with a cloth moistened with alcohol.
Changing the battery
In the course of time, batteries gradually wear out and require
longer charging times; this is normal. If you charge the batteries
regularly and notice a shortening of the operating time or an
extension of the charging time, it is probably time to replace them.
The new Li-Pol battery can be inserted simply by opening the lid
of the mPack UNPLUGGED:
• Isolate the mPack by removing it from the charger cradle or by
unplugging the UNPLUGGED mains transformer.
• Unscrew the eight screws in the lid of the mPack UNPLUGGED
with the screwdriver supplied and remove the lid, which contains
the Li-Pol battery.
• Make sure that no metal objects such as screws fall into the unit.
• Protect the electronics from electrostatic discharge (ESD), by
avoiding contact with electronic components.
• Before replacing the lid, remove the protective cap from the
three-pole battery contacts.
• Re-attach the lid with the eight screws, making sure that no wires
are damaged when doing so.
Note: The battery should be disposed of correctly and should be
completely discharged before disposal.
Note:
• Do not short-circuit the Li-Pol battery. This can occur when a
metal object such as a coin or paper clip connects the two
battery contacts together. This can damage the battery.
• Do not expose to fire (Explosion hazard).
10
Product specification
Technical Data
HEINE
EN50 UNPLUGGED:
®
Mains voltage:
100 – 240 V~ / 50 – 60 Hz
Current:
700 mA
Output:
9 VDC; 3.0 A
Fuse:
Integral overload protection
Duration of use:
Continuous
Classification:
Type BF
HEINE
mPack UNPLUGGED:
®
Input:
9 VDC; 1,5 A
Output:
Variable, max. 6.5 V / 10 W
Fuse:
Integral overload protection
Weight:
approx. 190 g
Dimensions:
162 mm x 52 mm x 24 mm
Duration of use:
depends on bulb type and brightness,
e.g. 5W bulb / HEINE OMEGA
approx. 120 min. at maximum brightness)
Charging time:
approx. 2 hours
HEINE Li-Pol Rechargeable battery:
Cell type:
Li-polymer
Nominal voltage:
7,4 V
Capacity:
approx. 1850 mAh
Environmental conditions:
Operation:
+10 °C to +40 °C
Storage:
+5 °C to +45 °C
Transport:
–20 °C to +45 °C
Rel. Humidity:
10% to 75%
Atmospheric pressure: 700 hPa to 1060 hPa
Applicable Standards
ISO 7000: 2004
(Graphical symbols for use on equipment)
EN ISO 15223-1
(Grafische Symbole zur Kennzeichnung)
IEC 60601-1: 1996 (Medical Electrical Equipment – PART 1)
EN 60601
UL 60601-1: 2003 (General requirements for Safety for Medical
Electrical Equipment)
IEC 60601-1-2
(Medical Electrical Equipment – COLLATERAL
STANDARD: ELECTROMAGNETIC
COMPATIBILITY – REQUIREMENTS AND
TESTS)
500,
®