Maintenance And Cleaning; Technical Characteristics - Mestra ANETO R-08051 Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

English
POWER SUPPLY BREAKDOWN
If power supply is interrupted (or main switch is
disconnected) while the unit is working, once the
power is recovered, the electronic control will try to
continue the program at the point it was stopped.
In order to correctly stop the unit, keep pressing the
START/STOP button (C) for some seconds until
pilots
and
blink, before disconneting the main
switch.
BLOCKED MOTOR
If the electronic control detects detects that the
motor is blocked (by example by solid gel), the
display will show the error message "cte".
The user will have to choose between two options:
1. Keep pressing the START/STOP button (C) to
end the cycle and solve the problem
2. Brievly press the START/STOP button (C) to
allow the unit heat the gel for 10 minutes befor
trying to start the motor again.

MAINTENANCE AND CLEANING

In order to guarantee optimum reliability and long
working life of the unit, we recommend you bear in
mind the following:
 Do not use solvents or any other inflammable
product.
 The unit is provided with a thermoprobe. If these
were to cause any problems, immediately
contact the MESTRA technical service.
 The motor reducer which drives the blades of
the stirrer is protected by a thermal fuse and by
the
microprocessor's
software, which disconnects the motor whenever
overloads occur.
 All the units are checked before leaving the
factory. During the rigorous final quality control
carried out in TALLERES MESTRAITUA, S.L.
several test are made using gel, in order to
better simulate the true use conditions of the
unit. For this reason, it is completely normal to
receive your unit with some remains of gel inside
the container.
PRECAUTIONS
 Before carrying out any maintenance on the
unit, switch it off and unplug from the power
supply.
 Before connecting the unit make sure the
electric connection is at 230 V, 50/60 Hz with
an earth connection.
 Neither children nor unqualified personnel are
allowed to manipulate the unit.
10
own
programming
 Do not open the upper lid when the unit is in
operation. Under circumstances must you
bridge the safety switch which protects the
opening of the lid.
 Avoid the level of gel rising above 3 cm over the
last blade (the closest to the upper edge of the
container).
 Place the unit upon a flat surface, correctly
levelled, far from any sources of vibration.
Avoid the possibility of it tipping over.
 Make sure the tap is turned off before starting
up the unit.
 Do not pour the gel at a temperature (C1),
because it may scald you. Wait until the pouring
temperatures reaches (C2).
 During operation avoid touching the lid. It may
have a high temperature. In the gel pouring
manoeuvre, wear gloves for protection.
 Be extremely careful during the start ups after a
motor blocking or power supply breakdown.
Eventual spattering or spillage of gel may
occur.
 Never leave the gel in liquid state during a
delayed start up. It may solidify, preventing a
normal start up of the unit. Always insert the gel
in pieces when programming a delayed start
up.

TECHNICAL CHARACTERISTICS

Voltage:
230 V, 50/60 Hz
Power:
650 W
Height:
470 mm
Width:
330 mm
Depth:
380 mm
Weight:
14 kg