4. Filters
a. Ensure filters are securely seated.
b. Perform the ULPA filter test (Section 6.2) prior to the start of each surgical procedure.
5. Tubing
a. Check throughout the surgical procedure to be sure the hose or tubing is not occluded.
Caution: Ensure sponge guard is attached to prevent surgical materials or tissue from
being aspirated / pulled into the tubing.
5.4
Clean-Up
After each surgical procedure, the SurgiFresh
cleaned and prepared for future use.
1.
Turn off the power switch and unplug the unit from the wall receptacle.
2.
Remove the unit from the immediate procedure area.
3.
Wipe the footswitch and footswitch cord with an appropriate hospital disinfectant. Coil the
footswitch cord around the cord wrap at the back of the unit.
4.
Wipe the power cord with an appropriate hospital disinfectant. Coil the power cord around
the cord wrap.
5.
Remove the tubing from the pre-filter or suction canister.
6.
If a suction canister was used, securely close lids/caps and properly dispose canister.
7.
Remove the pre-filter.
8.
Dispose of all single-use waste materials according to your institution's procedures for
biohazardous materials.
9.
Check the ULPA filter to be sure no fluids are on top of the filter. Replace filter if fluids are
present.
10.
Install a new pre-filter.
11.
Thoroughly wipe all external surfaces of the SurgiFresh
hospital disinfectant. Follow the procedures approved by your institution.
Warning: Ensure the unit is completely dry before restoring power.
6.0 MAINTENANCE
6.1 Power Cord
1.
Prior to each use, check the power cord along its entire length and at both plugs to ensure
no damage has occurred.
2.
Do not use a power cord with exposed wires, cracks, or frayed areas.
6.2 Filter Test
Note: Perform the ULPA filter test with no pre-filter or tubing in place.
1.
Ensure an ULPA filter is in the unit and seated properly.
2.
Plug the unit into an appropriate electrical receptacle.
OPERATING INSTRUCTIONS
MAINTENANCE
®
Turbo smoke evacuation system should be
®
8
Turbo with an appropriate