Compatible Sterilization Method: STERRAD™ in STERRAD sterilizer. Double
wrap with sterilization wrap. The full cycle of the Sterrad sterilizer comprises two
identical exposure phases (6 minute
plasma) preceded by a 10-minute pre-exposure plasma phase.
Follow the instructions provided with the STERRAD Sterilization System for
sterilization processing via this equipment and method.
NOTE: The LED Rover has a black anodized exterior finish which is affected by
Sterrad sterilization. The black anodized finish gradually bleaches; however,
this is irrelevant to the functionality of the unit.
7.3
HIGH LEVEL DISINFECTION (IMMERSION)
The LED Rover may be immersed in high level disinfectant solution no deeper than six
(6) inches (15cm) for as long as one (1) hour. Process according to the disinfectant
manufacturer's instructions or your institution's established protocol. We make no
claims other than the Rover's immersibility as stated.
8.
LIMITED WARRANTY
Your fiberoptic products(s) carry a 1-year warranty from the date of shipment on
workmanship and all defects of material except batteries. Should your product prove to
have such defects within one year from the date of shipment, ST TECHNOLOGIES will
repair or replace the product or component part without charge. Should your fiberoptic
product(s) need servicing under this warranty, please contact ST TECHNOLOGIES
return authorization documentation. You should carefully pack product in a sturdy carton
and ship it to the factory.
telephone number and a return address. Warranty does not cover equipment subject to
misuse, accidental damage, and normal wear and tear. This warranty gives you specific
legal rights and you may also have other rights that vary from state to state.
• In the U.S call:
• Customer Service
LIT102 CUDA
®
(English)
Please include a note describing the defects, your name,
Surgical
injection, 2 minute diffusion and 2 minute
877 814 2237
(toll free)
904 208 2291
Rev. O
Date of Revision:07/22/14
Page 7 of 40
for