The re-sterilization of reusable medical devices must be performed by a properly trained and
protected member of staff, in accordance with the regulations currently in force.
The re-sterilization protocol must take into account the risk of infection.
Refer to the manufacturer's instructions for each product used. Take particular care to adhere to the
instructions regarding concentrations, lengths of exposure, the replacement of solutions, and the
lifespan of the products. Never combine products. Adhere to the guidelines regarding the disposal
of used products.
Wear suitable protective clothing.
In order to avoid any risk of infection and injury, it is vital to wear protective gloves.
Never immerse the torque wrench in any kind of solution.
Never clean the torque wrench in an ultrasonic tank.
6.1. Cleaning
The device must be cleaned immediatly after the operation.
Use only detergent or disinfectant solutions with a neutral or slightly alkaline PH. It is extremely
unadvisable to use products which may fix X proteins to the unit (eg.alcohol, aldehydes...).
Do not use sodium hypochlorite (bleach) : there is a high risk of corrosion.
External cleaning
The torque wrench can be cleaned by using a brush under tap water. External disinfection using a
spray or a wipe soaked with disinfectant products microbiologically controlled (ph between 2.5-9).
In any case, it is important to follow the reaction time of the disinfectant product used.
Do not use disinfectant containing alkaline and chlorine.
Do not immerse the torque wrench in a disinfectant solution or in an ultrasound
container.