Cleaning
All components of the system can be both mechanically cleaned with detergent and water to
remove any visible contamination and chemically disinfected with chlorine.
The following represents guidance on the correct cleaning and decontamination process but does
not replace local policy and guidelines.
Toto Lateral Turning System Platform Cover
Keeping a cleaning record is recommended as the product is suitable for re-use.
Light Soiling
Clean with detergent and water to remove any visible signs of dirt and contamination and
decontaminate with a 0.1% chlorine-based solution (1,000ppm).
Rinse with clean water and wipe down with a single use, non-abrasive cloth and thoroughly dry
before reuse or storage.
Prolonged used of alcohol-based cleaning agents may reduce the life of the product. If used,
thoroughly rinse with clean water and allow to dry before use.
Heavy Soiling
In cases of gross contamination, clean with detergent and water to remove visible signs of dirt
and contamination and then decontaminate with a 1% chlorine-based solution (10,000ppm).
Do not apply a 1% chlorine-based solution for more than two minutes, doing so may damage the
cover long term.
Thoroughly rinse with clean water and wipe down with a single use, non-abrasive cloth and
thoroughly dry before reuse or storage.
Machine Washing Instructions
Cleaning and disinfection may be achieved by laundering at temperatures not exceeding 65ºC
for 10 minutes or 73ºC for three minutes .
Drying
To avoid shrinkage, line dry in a clean indoor environment. Thoroughly dry before re-fitting to the
Toto platform.
Do not mangle or iron.
Platform, Cabling, Control Unit and Air Hoses
Before cleaning, disconnect from the mains power supply.
The platform, cabling, control unit and air hoses can be wiped clean with alcohol wipes and a
chlorine derivative. Take care when wiping the control unit not to allow liquids to enter through
any apertures.
20