Radiated RF
Test
IEC61000-4-3
Frequen-
(Test speci-
cy
fications for
(MHz)
ENCLOSURE
PORT IMMUNITY
to RF wireless
5240
communications
equipment)
5240
5785
NOTE If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna and the
ME EQUIPMENT or ME SYSTEM may be reduced to 1 m. The 1 m test distance is permitted by IEC 61000-4-3.
a) For some services, only the uplink frequencies are included.
b) The carrier shall be modulated using a 50 % duty cycle square wave signal.
c) As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be used because while it does not
represent actual modulation, it would be worst case.
The MANUFACTURER should consider reducing the minimum separation distance, based on
RISK MANAGEMENT and using higher IMMUNITY TEST LEVELS that are appropriate for the reduced minimum
separation distance. Minimum separation distances for higher IMMUNITY TEST LEVELS shall be calculated using
the following equation:
E=6/d √P
Where P is the maximum power in W, d is the minimum separation distance in m, and E is the
IMMUNITY TEST LEVEL in V/m.
Band a)
Service
Modulation b)
(MHz)
a)
5 100 –
WLAN
Pulse
5 800
802.11
modulation b)
a/n
217 Hz
Modula-
Distance
tion b)
(m)
(W)
0,2
0,3
IMMUNITY
TEST LEVEL
(V/m)
9
27