CHARACTERISTICS
Size
Weight
Vacuum Pump
Variable Regulator (7)
Gauge (8)
Power Source: Battery (DC Power)
Medical switcher (AC Power)
Switch (5)
AC Receptacle (1A)
Collection Canister (13)
Patient tubing
Suction Tip
PROBLEM
Does not function when
switch is on (DC Power)
Blinking LED
Power source connected
No suction when pump
is running
General Specifications
S-SCORT
newDuet
®
Trouble Shooting
S-SCORT
®
newDuet
POSSIBLE CAUSE
Battery discharged
Low battery capacity
Fuse blown
AC power cord disconnected (1-1A)
Vacuum line (10) loose
Canister (13) defective
Canister lid (13) loose
Catheter thumb vent is open
Debris has collected in pump
8
Model 2314
®
SPECIFICATIONS
17"L x 9"H x 6"W
8.6 lbs.
12V DC oil-less diaphragm, 3.0A
Exceeds 525 mmHg
Clinical Airflow = 30LPM
Controls vacuum from 50 to 525+ mmHg
Calibrated in mmHg. Color-coded
Rechargeable Sealed Lead Acid 45 (±10%)
minute running time
100V - 240VAC, 47 - 63Hz
Uses 1ea. 2A 250V fast acting fuse
Off/Standby/On membrane switch
Uses (2) 2.5A 250 V fast acting fuses
1200cc SSCOR part #48041
Vinyl tubing 9/32"ID, 72"L SSCOR part
#43200
HI-D
®
. "Big Stick
®
". SSCOR part #44241
®
Model 2314
CORRECTIVE ACTION
Reconnect to AC power
to activate charger
Replace battery
Replace fuse (see above)
indicator light not lit
Reconnect cord
Tighten
Replace
Tighten
Occlude
Refer to maintenance
section of this manual