Cleaning the Autoclavable Remote Control
The remote control must be cleaned before sterilization.
Use a brush and only a mild, pH-neutral detergent to clean the remote con-
trol.
Make sure that there are no secretion residues, tissue residues or other fo-
reign bodies on the remote control before sterilization.
2
Disinfecting the Autoclavable Remote Control
Use a ph-neutral or slightly alkaline disinfectant. Comply with the instruc-
tions provided by the manufacturer of the disinfectant.
3
Sterilizing the Autoclavable Remote Control
The remote control was validated by the manufacturer for 100 sterilizati-
ons*. Never exceed the number of sterilizations specified by the manufac-
turer.
Only a cleaned, disinfected and dry remote control can be sterilized with
saturated steam. Please read the instructions in the operating manual for
your autoclave.
4
Sterilization conditions for the autoclavable remote
control
These instructions do not guarantee the sterility of the product. Your own
hygiene institute remains responsible for the validation of the sterilizati-
on and sterility.
The following sterilization methods has been validated by the manufactu-
rer:
Pre-Vac
Pre-vacuumed sterilizer
Wrapped or Unwrapped
5 min (273 °F/134 °C, 3 bar)
10 min
Gravity
Gravity-displacement sterilizer
Wrapped or Unwrapped
5 min (273 °F/134 °C, 3 bar)
10 min