Replacement (continued)
Install the Reagent
1.
Carousel Cover
2.
Prepare for Operation
1.
2.
3.
Verification
Perform Maintenance
1.
Procedure 3520
Temperature Status
2.
©1998 by Abbott Laboratories
Abbott Park, IL 60064
Install the reagent carousel cover.
Insert and tighten the three (3) screws
(non-captive). [1]
Close the processing center cover(s).
Power ON the processing module
Note
The system control center power
must be ON prior to turning on the
processing module to ensure proper
initialization.
To change the status of the processing module
and sample handler from Stopped to Ready,
perform the Startup procedure.
Select the As Needed category from the
maintenance screen under the system icon.
Perform diagnostic procedure 3520
Temperature Status.
RR-D2.04 SENSOR, TEMP, REFRIGERATOR
ABBOTT ARCHITECT™ i System Removal & Replacement Procedure
For information on the Startup procedure, refer to the ARCHITECT i
System Operations Manual, Section 5: Operating Instructions,
Subsection: Startup and Shutdown.
For further information on performing maintenance procedures, refer
to the ARCHITECT™ i System Operations Manual, Section 9: Service
and Maintenance, Subsection: As Needed Maintenance.
3 - 3
1
AMA-3183
ABBOTT ARCHITECT™ is a
trademark of Abbott Laboratories.
78154-102