Operating Instructions Continued
Canister Disposal After Use: The canister is for single use only. Replace the cap to the tip of the catheter (6) while the
pump is still running to trap fluids before they spill. Dispose of the canister according to local / regional / national
requirements for the disposal of hazardous waste. Install a new non-sterile single use canister on the unit.
Canister Installation and Storage: Slide the canister into the front rails of the chassis until the canister is locked in place.
Storing Canisters on the Chassis for Space Saving Portability:
Reverse the position of the canister and slide the body of the canister into the rails and the catheter tip through the
retention ring. Push the canister toward the chassis until the canister latch on top of the unit snaps into the detent on the
canister. To release the canister, pull the canister latch up and slide the body of the canister out.
WARNING: Periodic checking or replacement of the alkaline batteries is required. If for any reason poor battery quality is
suspected, perform a battery test as described below:
BATTERY TEST: Run the following test whenever poor battery quality is suspected. Turn the Quickdraw on
and run the unit for 30 seconds. Observe the battery condition indicator LED (2). If it lights red, it is time to
replace the battery (SSCOR part # 80613-100) or install 10 fresh AAA batteries into the SSCOR 80615
10xAAA Battery Holder. Dispose of the battery according to local / regional / national requirements for the
disposal of electronic waste.
Operating Temperature Range:
Operating Relative Humidity:
Operating Atmospheric Pressure:
Operating Altitude:
Storage & Transport Temperature:
Storage & Transport Relative Humidity:
Storage & Transport Atmospheric Pressure:
Transient Operating Temperature Range:
Transient Humidity Range:
Time to warm from minimum storage temperature to operating temperature: 30 minutes
Time to cool from maximum storage temperature to operating temperature: 30 minutes
6

Environmental Conditions

0 °C (+32 °F) – +40 °C (+104 °F)
0 - 93% (non condensing)
8.9 Psi (62 kPA) – 15.4 Psi (106 kPA)
<2000m
-40 °C (-40 °F ) – +50 °C (+122 °F)
0 - 95% (non condensing)
7.3 Psi (50 kPA) – 15.4 Psi (106 kPA)
-20 °C (-4 °F ) – +50 °C (+122 °F)
15% - 93% (non condensing)
Quickdraw - Model 2403 Series
Rev. E