Bowa COMFORT BOX Manual De Instrucciones página 14

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 35
4.
Detecting and rectifying errors
Message
No
COMFORT
instrument
detected
More than
one
COMFORT
instrument
detected
Cannot
read tag
6
Cause
Defective
instrument
memory
No BOWA
COMFORT
instrument
connected
Spring cover
compressed
Two instruments
simultaneously
connected
Defective
instrument
memory
Internal fault
COMFORT BOX
Remedy
Reconnect the
instrument.
If this message appears
again, using different
functional BOWA
COMFORT instruments,
please contact the
Technical Support.
1. Remove one instrument.
2.
If this message appears
again, please contact the
Technical Support.
Check the function using
another COMFORT
instrument.
If this message appears
again, using different
functional BOWA
COMFORT instruments,
please contact the
Technical Support.
11443-S0 EN

Publicidad

Tabla de contenido
loading

Productos relacionados para Bowa COMFORT BOX

Tabla de contenido