Connections; Calibration - Refco RESCALE Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

EN

6.3 CONNECTIONS

With the REFSCALE it is possible to establish a connection to the REFCO
REFMESH App or to the REFMATE digital manifold. To connect to REFMESH,
the blue connection LED on the remote control box must flashing, which
indicates that the scale is searching for a connection. As soon as a connection
is established, the LED lights up continuously.
Connection to REFMESH
With the REFMESH you can easily establish a connection with the REFSCALE
via the usual „connect" window on the REFMESH. Via REFMESH it is possible
to set tare, use the hold function, set an alarm and save a value which can be
shown in a report.
Connection to REFMATE
When the connection LED flashes you can easily search for a device with
the REFMATE. If the REFSCALE is within the range, the device connects
automatically. When the devices are connected, you can see the REFSCALE
values on the REFMATE. On the REFMATE you have access to the additional
screen by pressing the SWITCH button.

7.0 CALIBRATION

The REFSCALE is calibrated at the factory to provide high accuracy.
While recalibration should seldom, if ever, be necessary, recalibration is
performed easily without sending the product to the manufacturer for service.
A known calibration weight (in kilograms) is required to recalibrate the scale,
and can be any value between 10 kg and 100 kg. Best accuracy is obtained
by using the largest weight available.
To recalibrate the REFSCALE, perform the following steps:
1. Place the REFSCALE weighing platform on a rm, level surface.
2. Turn the power to the REFSCALE off.
3. Remove all objects from the platform.
REFSCALE Instruction Manual