Changing The Lamp; Changing The Fuse - 3B SCIENTIFIC PHYSICS 500 Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 17
dark between these positions. In contrast, ob-
jects that are isotropic and not birefringent re-
main dark in all positions.
4.3 Changing the lamp and fuse

4.3.1 Changing the lamp

Turn off the power switch, unplug the mains
plug and let the lamp cool down to avoid be-
ing burnt.
For safety reasons, remove the eyepiece.
To change the lamp lay the microscope on
its back to reach the lid on the bottom side.
Loosen screw C of the lamp socket and
push it outwards so that it is in the position
shown in Fig.1.
Loosen screw A and open the cover.
To remove the halogen lamp, use a cloth or
similar material. Do not touch the bulb with
the bare hand.
Lift out the halogen lamp and replace it with
a new one.
Close the cover and secure it with the
screw.
Push the lamp socket back into the original
position and tighten screw C.
C
B
A
Fig. 1 Lamp socket cover: A - knurled screw, B - ven-
tilation slots, C - securing screws of lamp-holder

4.3.2 Changing the fuse

Turn off the power switch and unplug the
mains plug.
Unscrew the fuse holder on the back of the
stand base with a screwdriver.
Replace the fuse and reinsert the holder in
its socket.
3B Scientific GmbH • Rudorffweg 8 • 21031 Hamburg • Germany • www.3bscientific.com
Subject to technical amendments
© Copyright 2013 3B Scientific GmbH
5. Storage, cleaning and disposal
Keep the microscope in a clean, dry and
dust free place.
When not in use always cover the micro-
scope with the dust cover.
Do not expose it to temperatures below 0°C
and above 40°C and a max. relative humid-
ity of over 85%.
Always unplug the mains plug before clean-
ing or maintenance.
Do not clean the unit with volatile solvents or
abrasive cleaners.
Do not disassemble objective or eyepieces
to attempt to clean them.
Use a soft linen cloth and some ethanol to
clean the microscope.
Use a soft lens tissue to clean the optics.
The packaging should be disposed of at
local recycling points.
Should you need to
dispose of the equip-
ment itself, never throw
it away in normal do-
mestic
waste.
Local
regulations for the dis-
posal
of
electrical
equipment will apply.

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

1003278

Tabla de contenido